由于 t:connect 应用程序的软件问题,200 多名糖尿病患者因胰岛素泵关闭而受伤。 200+ people with diabetes injured by insulin pump shutdowns due to software issue in t:connect app.
由于联网移动应用程序 t:connect 出现软件问题,导致胰岛素泵关闭,200 多名糖尿病患者受伤。 200+ people with diabetes injured by insulin pump shutdowns due to software issue in connected mobile app t:connect. 美国 FDA 发布了针对 Apple iOS t:connect 应用程序 v2.7 的一级召回,该应用程序与带有 Control-IQ 的 t:slim X2 胰岛素泵一起使用。 The US FDA issued a Class I recall for v2.7 of the Apple iOS t:connect app, used with t:slim X2 insulin pump with Control-IQ. 软件故障导致应用程序崩溃并重新启动,耗尽泵电池并导致胰岛素输送暂停,可能导致高血糖或糖尿病酮症酸中毒。 Software glitch causes app to crash and relaunch, draining pump battery and leading to insulin delivery suspension, potentially causing hyperglycemia or diabetic ketoacidosis. 截至4月15日,已报告224人受伤,无人死亡。 As of Apr 15, 224 injuries reported, no deaths.